Study on Energy Distribution of the 6 MeV Electron Beam using Gaussian Approximation

가우시안 근사를 이용한 6 MeV 전자선의 에너지분포에 관한 연구

  • Lee, Jeong-Ok (Department of Radiotechnology, Wonkwang Health Science College) ;
  • Kim, Seung-Kon (Department of Physics, Chonbuk national University)
  • Published : 1999.12.30

Abstract

A Gaussian distribution was parametrized for the initial distribution of the electron beam emitted from a 6MeV medical linear accelerator. A percent depth dose was measured in a water phantom and the corresponding Monte Carlo calculations were performed starting from a Gaussian distribution for a range of standard deviations, ${\sigma}=0.1$, 0.15, 0.2, 0.25, and 0.3 with being the mean value for the Incident beam energy. When measurement and calculation were compared, the calculation with the Gaussian distribution for ${\sigma}=0.25$ turned out to agree best with the measurement. The results from the present work can be utilized as input energy data in planning an electron beam therapy with a Monte Carlo calculation.

Keywords