Abstract
In this paper, the electromagnetic immunity of electronic medical equipments was tested by employing a standard test method for medical equipments. In addition, the electromagnetic interference to the medical equipments caused by four different types of mobile phones such as portable transceiver, analog, digital, and PCS cellular phones, have been studied. In this study, 16 medical equipments of 11 different types were investigated. There are a patient monitor, an infusion pump, a neonatal incubator and so on. In the case of the EMS test, the interferences were detected for seven medical equipments of four different types(53.3%) such as the patient monitor. For the two portable transceivers, 12 medical equipments of eight different models (75%) were affected. All medical equipments were satisfied with electromagnetic immunity by analog, digital and PCS cellular phones. The experiment results show that the malfunction of the electronic medical equipments is dependent on the frequency range and the electric field strength of the mobile phones.
본 논문에서는 전자파 표준 내성 측정 방법과 휴대무전기, 아날로그, 디지털 및 PCS 휴대전화와 같은 4기종의 휴대용 무선 통신기기에 의한 각종 의료기기의 내성을 시험하고 휴대용 무선 통신기기의 사용시 발생하는 의료기기의 오동작 여부를 측정, 분석하였다. 시험대상 의료기기로는 환자감시장치, 수액펌프, 보육기 등을 비롯하여 총 11종 16대의 의료기기를 선정하였다. 측정결과 표준 내성측정의 경우 환자감시장치를 비롯한 4종 7대(53.3%)의 의료기기가 오동작을 일으켰다. 휴대용 무선 통신기기 중 휴대무전기에 의한 의료기기의 장해 실험에서는 8종 12대(75%)의 의료기기가 오동작을 일으켰으나 아날로그, 디지털 및 PCS 휴대전화의 경우에는 모든 의료기기들이 내성 기준을 만족함을 알 수 있었다. 시험결과 의료기기에서 발생하는 방사 전자파의 고조파와 휴대무전기의 사용주파수가 유사한 경우와 휴대무전기의 사용 거리에 따른 전계강도의 세기에 따라서 의료기기의 오동작이 발생함을 알 수 있었다.