인체 및 인형 팬톰에서 전신방사선조사시 열형광선량계(TLD)를 이용한 선량분포 균일성 확인
In vivo and in vitro Confirmation of Dose Homogeneity in Total Body Irradiation with Thermoluminescent Dosimeter
- 지의규 (서울대학교 의과대학 치료방사선과학교실) ;
- 박석원 (성균관대학교 의과대학 삼성서울병원 치료방사선과) ;
- 강위생 (서울대학교 의과대학 치료방사선과학교실) ;
- 김일한 (서울대학교 의학연구원 방사선의학연구소)
- Chie Eui Kyu (Department of Therapeutic Radiology, Seoul National University College of Medicine) ;
- Park Suk Won (Department of Radiation Oncology, Samsung Medical Center) ;
- Kang Wee-Saing (Department of Therapeutic Radiology, Seoul National University College of Medicine) ;
- Kim Il Han (Institute of Radiation Medicine, Medical Rearch Center, Seoul National University)
- 발행 : 1999.12.01
초록
목적 : 면역억제, 악성림프종 및 백혈병의 치료, 골수이식 등에 적용되는 전신방사선치료에 시행되는 좌우 이문대향조사법의 표면 방사선량 및 중심 방사선량의 균일성을 열형광선량계를 이용하여 확인하고자 하였다. 대상 및 방법 : 1996년 9월부터 1998년 8월까지 좌우 이문대향조사법으로 전신방사선치료를 시행한 10명의 환자 중 매번 방사선량이 측정된 8명의 환자를 대상으로 하였다. 열형광선량계를 사용하여 두부, 경부, 액와부, 대퇴부, 족관절부 등의 표면 선량을 측정하였고 인형 팬톰에서 동일한 조건하에서 두부, 경부, 액와부, 복부, 둔부 등의 표면 선량과 중심 선량을 측정하였다. 결과 : 중심부인 복부 처방 선량을 기준으로 할 때 부위별로 환자의 두부, 경부, 액와부, 대퇴부, 족관절부에서 표면 방사선량은 각각
Purpose : Total body irradiation (TBI) or whole body irradiation is used to acquire immune suppression, to treat malignant lymphoma and leukemia, and as an conditioning regimen for bone marrow transplantation. For these purposes, many methods were developed to obtain homogenous dose distribution. The objective of this study was to analyze and confirm the accuracy and the homogeneity of the treatment setup, the parallel opposed lateral technique, currently used in Seoul National University Hospital. Materials and Metheods : Surface dose data, measured with a thermoluminescent dosimeter, of 8 patients among 10 patients, who were given total body irradiation with the parallel opposed lateral technique between September 1996 to August 1998, at Seoul National University Hospital were analyzed. Surface doses were measured at the head, neck, axilla, thigh, and ankle level. Surface and midline doses were measured with similar set-up and technique in the Humanoid phantom. Results : Measured surface doses relative to prescribed dose for the head, neck, axilla, thight, and ankle leve were