Analysis of Image Factors of X-ray Films: Study for the Intelligent Replenishment System of Automatic Film Processor

자동현상기 지능화에 필요한 연산처리 기법의 개발을 위한 방사선 필름의 영상 지수의 분석

  • Park, Sung-Tae (Department of Diagnostic Radiology, Asan Medical Center University of Ulsan College of Medicine) ;
  • Yoon, Chong-Hyun (Department of Diagnostic Radiology, Asan Medical Center University of Ulsan College of Medicine) ;
  • Park, Kwang-Bo (Department of Diagnostic Radiology, Asan Medical Center University of Ulsan College of Medicine) ;
  • Auh, Yong-Ho (Department of Diagnostic Radiology, Asan Medical Center University of Ulsan College of Medicine) ;
  • Lee, Hyoung-Jin (Department of Diagnostic Radiology, Asan Medical Center University of Ulsan College of Medicine) ;
  • In, Kyung-Hwan (Department of Diagnostic Radiology, Asan Medical Center University of Ulsan College of Medicine) ;
  • Kim, Keon-Chung (Department of Diagnostic Radiology, Asan Medical Center University of Ulsan College of Medicine)
  • 박성태 (울산대학교 의과대학.서울중앙병원 진단방사선과) ;
  • 윤종현 (울산대학교 의과대학.서울중앙병원 진단방사선과) ;
  • 박광보 (울산대학교 의과대학.서울중앙병원 진단방사선과) ;
  • 오용호 (울산대학교 의과대학.서울중앙병원 진단방사선과) ;
  • 이형진 (울산대학교 의과대학.서울중앙병원 진단방사선과) ;
  • 인경환 (울산대학교 의과대학.서울중앙병원 진단방사선과) ;
  • 김건중 (울산대학교 의과대학.서울중앙병원 진단방사선과)
  • Published : 1998.06.30

Abstract

We analyzed image factors to determine the characteristic factors that need for intelligent replenishment system of the auto film processor. We processed the serial 300 sheets of radiographic films of chest phantom without replenishment of developing and fixation replenisher. We took the digital data by using film digitizer which scaned the films and automatically summed up the pixel values of the films. We analyzed characteristic curves, average gradients and relative speeds of individual film using densitometer and step densitometry. We also evaluated the pH of developer, fixer, and washer fluid with digital pH meter. Fixer residual rate and washing effect were measured by densitometer using the reagent methods. There was no significant reduction of the digital density numbers of the serial films without replenishment of developer and fixer. The average gradients were gradually decreased by 0.02 and relative speeds were also gradually decreased by 6.96% relative to initial standard step-densitometric measurement. The pHs of developer and fixer were reflected the inactivation of each fluid. The fixer residual rates and washing effects after processing each 25 sheets of films were in the normal range. We suggest that the digital data are not reliable due to limitation of the hardware and software of the film digitizer. We conclude that average gradient and relative speed which mean the film's contrast and sensitivity respectively are reliable factors for determining the need for the replenishment of the auto film processor. We need more study of simpler equations and programming for more intelligent replenishment system of the auto film processor.