AN EVALUATION OF PRECISION FIT OF IMPLANT-SUPPORTED PROSTHESIS USING THE PERIOTEST

$Periotest^{(R)}$를 이용한 임프란트지지 보철물의 적합도 평가에 관한 연구

  • Kim, Young-Min (Dept. of Prosthodontics, College of Dentistry, Chonnam National University) ;
  • Bae, Jeong-Sik (Dept. of Prosthodontics, College of Dentistry, Chonnam National University)
  • 김영민 (전남대학교 치과대학 보철학교실) ;
  • 배정식 (전남대학교 치과대학 보철학교실)
  • Published : 1998.08.01

Abstract

In this study, the Periotest value was measured with Periotest to evaluate precision fit of the 2-unit and 3-unit implant-supported prosthesis by modifying the size and location of ill-fitted conditions. The 2-unit prosthesis was fabricated with the right implant fitted incorrectly and the 3-unit prosthesis with the right and center implant fitted incorrectly. To evaluate the effects of the ill-fitted sizes, 4 groups were divided.:The control group being the accurately fabricated sample group fitted properly. Group 1 was constructed with $40{\mu}m$ ill-fitted conditions, group 2 with $70{\mu}m$ and group 3 with $100{\mu}m$ ill-fitted conditions. The Periotest value was measured at each implant site after tightening 10Ncm. The result was follows : 1. The PTV on the ill-fitted area in the 2-unit implant-supported prosthesis increased as the ill-fitted conditions increased. There was a statistically significant difference among groups(p<0.05). In the same ill-fitted sample, the PTV depending on the measured location demonstrated a statistically significant difference (p<0.05) 2. The PTV on the ill-fitted area of the 3-unit implant-supported with an ill-fitted condition in the right implant increased as the ill-fitted conditions increased. There was a significant difference among groups (p<0.05). In the same ill-fitted sample, the PTV depending on the measured location demonstrated a statistically significant difference (p<0.05). 3. In the 3-unit implant-supported prosthesis with ill fitting conditions in the center implant, the PTV on the ill-fitted area demonstrated a statistically significant difference between the control group, group 1 and group 2 (p<0.05). In the same ill-fitted sample, the PTV depending on the measured location demonstrated significant difference between the gap side and the adjacent side with over $70{\mu}m$ ill-fitted conditions (p<0.05). The results suggest that Periotest is a valuable objective method for evaluating the precision fit of an implant superstructure.

Keywords