인공치근용 CP-Ti과 Ti 합금의 기계적 성질 및 내식성

Mechanical Properties and Corrosion Resistance of CP-Ti and Ti Alloy for Dental Implants

  • 김연욱 (계명대학교 공과대학 재료공학과) ;
  • 정종평 (서울대학교 치과대학)
  • 발행 : 1998.10.20

초록

Commercially pure titanium(cp-Ti) and Ti-15wt%Zr-4wt%Nb-4wt%Ta alloy were melted in vacuum induction furnace. According to the chemical analysis, the content of carbon was above ASTM standard in the cast ingots because of using graphite crucible. The TEM micostructures of cp-Ti and Ti alloy shows that chemically stable TiC precipitates distribute in ${\alpha}-Ti$ matrix. In order to examine the properties of cp-Ti and Ti-Zr-Nb-Ta alloy for dental applications, mechanical properties and corrosion resistance were investigated. The anodic polarization properties of Ti-Zr-Nb-Zr alloy were almost same as that of cp-Ti in 1% lactic acid. However, as the results of the anodic polarization test in 5% HCl, it was known that Ti-Zr-Nb-Zr alloy showed a rapid decrease in current density at higher potential in comparison with cp-Ti. The yield stress and tensile strengh in Ti-Zr-Nb-Ta were ${\sigma}_{0.2}=623\;MPa$, ${\sigma}_{T.S.}=708\;MPa$ and these results showed 30% increase in yield stress in comparison with cp-Ti.

키워드