The Development of a Cryotherapy System
한냉물리치료기의 개발
- Kim, Yeong-Ho (Dept. of Medical Engineering, Research Institute of Medical Engineering, College of Health Science, Yonsei University) ;
- Yang, Gil-Tae (Korea Orthopedics and Rehabilitation Engineering Center) ;
- Jang, Yun-Hui (Korea Orthopedics and Rehabilitation Engineering Center) ;
- Park, Si-Bok (Department of Rehabilitation Medicine, Hanyang University Hospital) ;
- Ryu, Jin-Sang (Century Co.)
- 김영호 (연세대학교 보건과학대학 의용전자공학과, 의공연구소) ;
- 양길태 (의공학연구소 재활공학연구센터) ;
- 장윤희 (의공학연구소 재활공학연구센터) ;
- 박시복 (한양대학교병원 재활의학과) ;
- 류진상 (주식회사 센추리)
- Published : 1998.12.01
Abstract
A cryotherapy system using cold air was developed. The developed system had superior low-temperature characteristics with various flow rates and nozzle sizes, and used R-404A, as a coolant, which has no destructive effects of Ozone layers. Flow rates and the treatment time can be easily altered during the operation. In addition, and alarm system was designed for the overload, overheat, and over-charge of the machine. For clinical applications, skin temperatures, intra-articular temperatures of the knee joint and intra-muscluar temperatures of the gluteal muscles were measured during and after the cryotherapy. After a 5-minute therapy, skin and intra-articular temperatures decreased by
신냉매인 R-404A를 사용하여 저온특성이 우수한 한냉물리치료기를 개발하였다. 임상평가를 통해서 침온도계를 슬관절 강 내 및 둔부근육 내에 삽입하여 직접 근육온도를 측정하였고, 적외선체열촬영기를 이용하여 표피온도를 객관적으로 측정하였다. 슬관절 부위에 대한 5분 동안의 한냉물리치료에 따른 표피 및 관절강 내 온도변화를 측정한 결과, 표피는 23.3