Reproducibility of the Assessment of Myocardial Function Using Gated Tc-99m-MIBI SPECT and Quantitative Software

게이트 Tc-99m-MIBI SPECT와 Cedars 소프트웨어를 이용한 심근 기능 측정의 재현성

  • Lee, Dong-Soo (Department of Nuclear Medicine, Seoul National University College of Medicine) ;
  • Cheon, Gi-Jeong (Department of Nuclear Medicine, Seoul National University College of Medicine) ;
  • Ahn, Ji-Young (Institute of Radiation Sciences, Seoul National University Medical Research Center) ;
  • Chung, June-Key (Department of Nuclear Medicine, Seoul National University College of Medicine) ;
  • Lee, Myung-Chul (Department of Nuclear Medicine, Seoul National University College of Medicine)
  • 이동수 (서울대학교의과대학 핵의학교실) ;
  • 천기정 (서울대학교의과대학 핵의학교실) ;
  • 안지영 (서울대학교의과대학 방사선의학연구소) ;
  • 정준기 (서울대학교의과대학 핵의학교실) ;
  • 이명철 (서울대학교의과대학 핵의학교실)
  • Published : 1998.10.30

Abstract

Purpose: We investigated reproducibility of the quantification of left ventricular volume and ejection fraction, and grading of myocardial wall motion and systolic thickening when we used gated myocardial SPECT and Cedars quantification software. Materials and Methods: We performed gated myocardial SPECT in 33 consecutive patients twice in the same position after Tc-99m-MIBI SPECT We used 16 frames per cycle for the gating of sequential Tc-99m-MIBI SPECT. After reconstruction, we used Cedars quantitative gated SPECT and calculated ventricular volume and ejection fraction (EF). Wall motion was graded using 5 point score. Wall thickening was graded using 4 point score. Coefficient of variation for re-examination of volume and fraction were calculated. Kappa values (k-value) for assessing reproducibility of wall motion or wall thickening were calculated. Results: Enddiastolic volumes (EDV) ranged from 58 ml to 248 ml (122 ml +/- 42 ml), endsystolic volumes (ESV) from 20 ml to 174 ml (65 ml +/- 39 ml), and EF from 20% to 65% (51% +/- 14%). Geometric mean of standard deviations of 33 patients was 5.0 ml for EDV, 3.9 ml for ESV and 1.9% for EF. Their average differences were not different from zero (p>0.05). k-value for wall motion using 2 consecutive images was 0.76 (confidence interval: 0.71-0.81). k-value was 0.87 (confidence interval: 0.83-0.90) for assessment of wall thickening. Conclusion: We concluded that quantification of functional indices, assessment of wall motion and wall thickening using gated Tc-99m-MIBI SPECT was reproducible and we could use this method for the evaluation of short-acting drug effect.

목적: 좌심실의 부피와 구혈률을 Cedars 소프트웨어로 구하였을 때 정량한 값과 심근벽운동과 심근의 수축기 두꺼워짐을 등급으로 평가한 방법의 재현성을 알아보고자 하였다. 대상 및 방법: 33명의 무작위 추출된 환자를 상대로 통상의 이중동위원소 휴식/부하 심근관류 SPECT 때 부하 Tc-99m-MIBI 게이트 SPECT를 촬영한 후 이어서 같은 자리에서 게이트 SPECT를 한번 더 촬영하였다. 재구성한 후 Cedars 소프트웨어로 확장기말 부피와 수축기말 부피, 구혈률을 측정하여 연속 측정의 재현성을 평균변이계수와 Bland Altman 도표를 그려 분석하였다. 벽운동을 5등급으로 점수를 매겨 연속촬영한 영상의 판독 재현성을 조사하였다. 심근벽의 수축기 두꺼워짐을 역시 4등급으로 점수를 매겨 판독 재현성을 조사하였다. 결과: 확장기말 부피와 수축기말부피는 평균변이계수가 5.0 ml, 3.9 ml이고 구혈률의 평균변이계수는 1.9%이었다. 2표준편차 범위는 확장기말 부피는 18 ml, 수축기말부피는 17 ml이었다. 심근벽운동과 심근의 수축기 두꺼워짐은 kappa 값이 0.7로 판독재현성이 우수하였다. 결론: 이 연구의 결과로 같은 환자를 같은 위치에서 두 번 연이어 게이트 심근 Tc-99m-MIBI SPECT를 촬영할 때, 우리가 산출한 변이가 우수하고 약물효과를 판정할 때 기준으로 삼을 오차범위를 어느 정도로 정해야 하는지 확립하였다.