Improvement of size measurement polystyrene spheres of diameters 3$\mu$m and 10$\mu$m by optical microscope with CCD camera

CCD 카메라가 장착된 광학현미경을 사용한 폴리스티렌구 (3 $\mu$m와 10 $\mu$m)의 평균지름측정

  • Published : 1998.12.01

Abstract

Center Distance Finding (CDF) is a technique to find the sphere diameter by measuring the distance between two contacting spheres. The focal spots of the sphere clusters are formed in the back-focal plane by the transmission-mode optical microscope with the pseudothermal illumination source. Digital images taken by the CCD camera were processed by the software called Global Lab Image. The centers of the focal spots are found and the spot positions are expressed in terms of the CCD pixel elements, whose coordinate are calibrated by a heterodyne interferometer. The new CDF measurement system has been developed, which are more advantageous in time and convenience than the existing system, while the measurement uncertainly remains sufficient for its use as a magnification standard for optical microscopy. Two kinds of polystyrene spheres whose nominal diameters 3 and 10 $\mu\textrm{m}$ (NIST SRM 1962 and 1960) are measured with the uncertainly less than 1% at the confidence level of 99%, and the results are compared with the results of National Institute of Standards and Technology.

중심거리측정법은 서로 붙어있는 두 입자 중심점간의 거리를 측정하여 입자의 지름을 구하는 방법이다. 표면장력에 의해 배열이 형성된 시편 입자들의 초점군을 투과식 광학현미경에 평행 레이저광을 입사시켜 얻어내고 CCD 카메라로 영상을 받아 전산 분석하였다. Global lab image라는 영상처리 프로그램으로 초점들의 중심점을 찾고 붙어있는 입자들의 중심점간 거리를 화소(CCD 카메라의 pixel)단위로 계산하였으며, 화소의 좌표는 레이저 간섭계로 변위를 읽는 이송대를 이용하여 교정하였다. 기존의 측정방법을 개선하여 빠른 시간에 간편하게 측정하면서도 표준입자의 배율고정에 충분한 불확도를 얻을 수 있었다. 본 실험에는 NIST 인증물질인 3$\mu\textrm{m}$와 10$\mu\textrm{m}$ 폴리스티렌구(NIST SRM 1962, 1960)를 측정하였으며, 1%이하의 불확도(신뢰도 99% 수준)로써 NIST 결과와비교하였다.

Keywords

References

  1. MD 20089 Particle size standard and their certification at NIST A. W. Hartman
  2. proceedings of the international symposium on optical paricle sizing: theory and practice Thomas R. Lettieri
  3. The International Conference on Liquid borne Particle inspection and Metrology Standards for particle size A. W. Hartman
  4. 전북대학교 석사학위청구 논문 Mie산란을 이용한 작은 유전체공의 평균지름 측정 손공숙(외)
  5. 한국광학회지 v.9 no.4 광산란을 이용한 0.3㎛ 폴리스티렌구의 평균지름측정 불확도 박병천(외)
  6. 한국표준과학연구원 2차년도 보고서 입도 표준확릭(KRISS-96-041-IR) 엄천일외
  7. JRNBS v.90 no.1 Development of a one-Micrometer-Diameter Particle Size Standard Reference Material G. W. Mulholland(et al.)
  8. Particulate science and Technology v.7 Improved Array Method for Size Calibration of Mondispherse Spherical Particles by optical Microscope S. D. duke(et al.)
  9. Journal of Research NIST v.96 no.6 Certification of SRM 1960 : 10 ㎛ Diameter Polystyrene Spheres Thomas R. Lettieri (et al.)
  10. Journal of Research NIST v.97 no.2 Certification of NIST SRM 1962 : 3 ㎛ Diameter Polystyrene Spheres Arie W. Hartman(et al.)
  11. J. Colll. Interf. Sci. v.53 B. V. Derjaguin;V. M. Muller;Yu P. Toporov
  12. Proc. Roy. Soc. Lond v.A324 K. L. Johnson;K. Kendall;A. S. Roberts