Journal of Radiation Protection and Research
- Volume 23 Issue 3
- /
- Pages.159-174
- /
- 1998
- /
- 2508-1888(pISSN)
- /
- 2466-2461(eISSN)
Development of TLD Algorithms by Monochromatic Fluorescence Radiations and Continuous Spectrum X-rays
단일에너지 형광 X선 및 연속 스펙트럼 X선장에 의한 TLD 알고리즘 개발
- Kim, Jang-Lyul (Korea Atomic Energy Research Institute) ;
- Kim, Bong-Hwan (Korea Atomic Energy Research Institute) ;
- Chang, Si-Young (Korea Atomic Energy Research Institute) ;
- Lee, Jai-Ki (Han-Yang University)
- Published : 1998.09.30
Abstract
Personal dosimetry system is required to measure the personal dose equivalent accurately in a wide range of radiation fields, but the dose evaluation algorithms have been developed with the X-ray fields described in MOST Ordinance (equivalent to the ANSI N13.11) from which the actual fields to be monitored may be significantly different. To evaluate the dose more accurately when workers are exposed to the non-ANSI N13.11 radiation fields, two algorithms for monochromatic radiations (one algorithm was used for various ratios of TL dosimeter and the other for matrix approximation) were developed with the experimental data of the energy responses of the
개인방사선측정시스템은 광범위한 에너지의 방사선에 대하여 정확한 개인선량당량을 측정할 수 있어야 하지만 현재 사용중인 선량평가 알고리즘은 ANSI N13.11의 방사선장과 거의 유사한 과학기술부 고시에 명시되어 있는 X-선장을 기준으로 개발되었고 이는 실제 측정이 요구되는 방사선장과 상당한 차이가 있을 수 있다. 방사선작업종사자가 ANSI N13.11의 방사선장과 다른 형태의 방사선에 피폭되었을 경우 좀더 정확한 선량을 평가하기 위하여 한국원자력연구소에 설치되어 있는 단일에너지 X-선장에
Keywords