Improved stress analyses of dental systems implant by homogenization technique

균질화기법을 이용한 치과 임플란트 시스템의 응력해석

  • Koh, Chul-Su (Department of Periodontology, College of Dentistry, Kyung-Hee University) ;
  • Lee, Man-Sup (Department of Periodontology, College of Dentistry, Kyung-Hee University) ;
  • Choi, Kui-Won (Korea Institute of Science and Technology)
  • 고철수 (경희대학교 치과대학 치주과학교실) ;
  • 이만섭 (경희대학교 치과대학 치주과학교실) ;
  • 최귀원 (한국과학기술연구원)
  • Published : 1997.03.30

Abstract

Homogenization technique is employed to investigate the series of stress analyses of mandible for three different types of dental implants. This technique helps to make proper material model of bone and analyze such a non homogeneous structure at the level of individual microstructural unit. The stress analyses with homogenization technique show much higher stress level in the sponge bone, compared to those of conventional FEM. It also manifested that even a minor lateral force results in crucial stresses in the dental implant system and that the macroscale model should take the shape and size after real mandible to produce reasonable solution in the analyses of dental implant systems. The shapes of dental implants simulated in this study are rectangular-cross-sectioned type, hemi-sphere rooted type, and wedge type implant. The stress states of mandible with hemisphere rooted type implant and wedge type implant show similar levels, while those with sectioned rectangular implant results in higher stresses. It is suggested that the distance between the implant tip and cortical bone be kept far enough to prevent stress concentrations in the mandible.

Keywords