Analytical Science and Technology (분석과학)
- Volume 10 Issue 1
- /
- Pages.35-42
- /
- 1997
- /
- 1225-0163(pISSN)
- /
- 2288-8985(eISSN)
Analytical method of trace elements in ceramic capacitor materials
세라믹 축전물질들의 미량성분 분석방법
Abstract
The sample pretreatment technique using a high pressure acid digestion bomb for the analysis of ceramic capacitor materials, such as barium titanate and PZT, was studied. When the concentrations of hydrochloric acid were varied with the addition of nitric acid or distilled water, quantitative measurements for those samples were carried out using inductively coupled plasma atomic emission spectrometry. From this experiment the results indicate that most of elements, such as Ba, Mn, Zn, Si, etc., aren't affected by the concentration of hydrochloric acid but Nb and Zr are very susceptable to it. It however turns out that the digestion time relatively gave little effects on the analytical result. In case of Nb the ratio of hydrochloric acid to water should be greater than 3:1(v / v) for the best analytical result. For the Pb determination use of diluted hydrochloric acid compared to the mixture of nitric acid and hydrochloric acid showed a better analytical result.
Barium titanate 및 PZT와 같은 세라믹 축전물질의 분석을 위해 고압 산분해법을 이용한 시료 전처리 기술을 연구하였다. 질산이나 증류수를 첨가하고 염산의 농도를 변화시키면서, 이들 시료의 정량분석을 유도결합 플라즈마 방출분광법에 의해 수행하였다. 분석 결과를 보면 Ba, Mn, Zn, Si 등 대부분의 원소들은 염산의 농도에 영향을 받지 않으나, Nb와 Zr은 이에 민감한 것으로 밝혀졌다. 그러나 분해시간은 상대적으로 분석 결과에 거의 영향을 미치지 않았다. Nb의 경우에는 가장 좋은 분석 결과를 얻기 위해 염산과 증류수의 비가 3:1(v / v) 보다는 더 진해야 함을 알았다. 또한 Pb의 분석에 있어서 염산과 질산의 혼산을 사용하는 경우보다 붉은 염산만을 사용하였을 경우에 더 좋은 분석 결과를 나타내었다.