Abstract
The shape of TL glow curve is a useful indicator for assurance of correct reading of the personal dosimeter. Since the reading procedure of TLD is irreversible, however, an analytic remedy should be considered to procure reliable dosimetric information for the readings with irregular glow con shape. In this study, kinetic trapping parameters of CaSO$_4$ : Dy Teflon personal dosimeter(Teledyne PB-6A) were analyzed by Halperin and Braner's model for general-order kinetics. From these kinetic tapping parameters, we also developed a simple procedure to retrieve the dosimetric information from abnormally distorted glow curves. The computerized glow curve deconvolution(CGCD) fitting of the reference glow curve with kinetic parameters from this study yields relative errors of about 5% from the expected integral. It was also found that the glow curve remedial procedure developed could retrieve the distorted TL glow curves within ewer ranges of 1575. With the glow curve retrieval techniques, doses incurred by gamma radiation can now be successfully re-constructed for the CaSO$_4$ : Dy Teflon dosimeter resulting abnormal glow curves.