Intercomparison of vacuum standards of Korea, United Kingdom, and Japan

진공표준의 국제비교 연구

  • 홍승수 (한국표준과학연구원 압력진공그룹) ;
  • 신용현 (한국표준과학연구원 압력진공그룹) ;
  • 임종연 (한국표준과학연구원 압 력진공그룹) ;
  • 이상균 (한국표준과학연구원 압력진공그룹) ;
  • 정광화 (한국표준과학연구 원 압력진공그룹)
  • Published : 1997.11.01

Abstract

TDS (Thermal Lkso~ption Spectroscopy)system, for diagnosis of CRT manufacturing process, was designed and constructed. Outgassings and themla1 desorptions from the part or materials of CRT can be measured and analysed with this system at various temperatures. The system is consisted of 3 pirrts. vacuum chamher and pumping system with variable conductance, sample heating stages & their controller, and outgassing measurement devices, like as ion gauge or quadrupole mass spectrometer. The ultimate pressure of the system was under $1\times10^{-7}$ Pa. With the variable conductance system, the effective pumping speed of the chamber could he controlled from sub 11s to 100 11s. The effective pumping speed values were determined by dynamic flow measurement principle. The temperatures and ramp rate of sample were controlled by tungsten heater and PID controller up to $600^{\circ}C$ within t $\pm 1^{\circ}C$$difference to setting value. Ion gauge & QMS were calibrated for quantitative measurements. Some examples of TDS measurement data ;ind application on the CRT process analysis were shown.

Capacitance Diaphragm Gauge(CDG)를 전달표준기로 사용하여 한국과 영국의 저진 공표준의 국제비교를 수행한 결과 양국 표준기에서 교정한 CDG의 압력비의 차이는 0.05% 이었다. 또한 두 개의 Spinning Rotor Gauge(SRG)를 이용하여 한국과 일본이 고진공표준의 국제비교를 수행한 결과 양국 표준기에서 교정한 SRG의 accommodation coefficient($sigma$)의 차이가 0.4% 이내이었다. CDG와 SRG를 이용한 한국과 영국의 저진공표준기 및 한국과 일 본의 고진공표준기의 국제비교 결과 CDG 및 SRG 자체의 고유오차범위 이내에서 결과가 일치함을 확인할 수 있었다.

Keywords