A pixel profile in the conventional DANTE sequence is so poor that the excited area by DANTE sequence is a small portion of a pixel. This causes poor signal to noise ratio in DANTE image. In this paper, a frequency modulated(FM) DANTE imaging sequence is proposed to improve pixel profile in DANTE image. A DANTE pulse train is shaped by an FM function so that all the spins within a pixel are excited, thereby improving the signal to noise ratio. It also shows that the pixel profiles are dependent on the sweep in FM signal. Computer simulations and experimental result obtained using a 7.0 T NMR imaging system are presented.

Keywords

References

  1. J. Magn. Reson v.29 Selective Excitation in Fourier Transform Nuclear Magnetic Resonance G. A. Morris;R. Freeman
  2. J. Magn. Reson v.23 Simple Pulse Sequence for Selective Excitation in Fourier Transform NMR G. Bodenhausen;R Freeman;G. A. Morris
  3. Magn. Reson in Med v.15 A DANTE Tagging Sequence for the Evaluation of Translational Sample Motion T. J. Mosher;M. Smith
  4. J. .Magn. Reson series B v.101 DANTE Ultra fast Imaging Sequence(DUFIS) I. J. Lowe;R. E. Wysong
  5. Books of Abstract, 7th SMRM Fast Imaging Burst Excitation Pulses J. Hennig
  6. J. Magn. Reson v.27 Planar Spin Imaging by NMR P. Mansfield;A. A. Maudsley
  7. IEEE Trans. Med. imaging. v.5 no.1 High-Speed Spiral-Scan Echo Planar NMR Imaging-Ⅰ C.B.Ahn;J.H.Kim;Z.H. Cho
  8. Books of Abstract, 6th SMRM v.2 Phase Error Corrected Interlaced Echo Planar Imaging Z.H.Cho;C.B.Ahn;J.H.Kim;Y.E.Lee;C.W.Mun
  9. Magn. Reson in Med v.3 RARE Imaging:A Fast imaging Method for Clinical MR J. Hennig;A. Nauerth;H. Friedburg
  10. Magn. Reson in Med v.3 Use of Frequency-Modulated Radiofrequency Pulses in MR Imaging Experiments D. Kunz
  11. Books of Abstract, 12th SMRM Optimized Ultra-fast Imaging sequence(OUFIS) L. Zha;I.J. Lowe
  12. Magn. Reson in Med v.33 Optimized Ultra-fast Imaging sequence(OUFIS) L. Zha;I.J. Lowe
  13. US patent 532243 Fringe Field NMR Z.H.Cho;E.Wong
  14. J. Appli. Phys v.26 Spin Echo Serial Storage Memory A. Anderson;R. Garwin;E. Hahn(et al.)
  15. J. Magn. Reson v.75 Shaped Selective Pulses for Coherence-Transfer Experiments J. Fridrich;S. Davis;R. Freeman
  16. J. Magn. Reson v.76 Dynamic Range Improvement in NMR Imaging Using Phase Scrambling A. A. Maudsley
  17. Books of Abstract, 10th SMRM Burst Excitation Pulses P. Leroux;J. Pauly
  18. Magn. Reson in Med v.28 A Novel Flow Suppression Technique using Tailored RF Pulses Y. M. Ro;Z. H. Cho