Comparison Study of Experimental Neutron Room Scattering Corrections with Theoretical Corrections in RCL's Calibration Facility at KAERI

한국원자력연구소 중성자교정실에 대한 중성자산란보정인자 결정연구

  • Published : 1997.03.30

Abstract

Neutron room scattering corrections that should be made when neutron detectors are calibrated with a $D_2O$ moderated $^{252}Cf$ neutron source in the center of a calibration room are considered. Such room scattering corrections are dependent on specific neutron source type, detector type, calibration distance, and calibration room configuration. Room scattering corrections for the responses of a thermoluminescence dosimeter and two different types of spherical detectors to neutron source in the Radiation Calibration Laboratory(RCL) neutron calibration facility at the Korea Atomic Energy Research Institute(KAERI) were experimentally determined and are presented. The measured room scattering results are then compared with theoretical results calculated by predicting room scattering effects in terms of parameters related to the specific configuration. Agreement between measured and calculated scattering correction is generally about 10% for three kinds of detectors in the calibration facility.

중성자교정실내에서 $D_2O$ 감속 $^{252}Cf$중성자선원을 사용하여 계측기를 교정할 때는 그 계측기에 대한 교정실산란보정 인자를 미리 결정하여야 한다. 이러한 교정실산란보정인자는 계측기의 종류, 교정거리, 교정실형태에 따라 다르게 결정된다. 본 연구에서는 한국원자력연구소에서 운영하는 2차 표준중성자교정실에서 한가지의 열형광선량계와 2가지의 구형검출기에 대한 교정실산란보정인자를 실험적으로 결정하였고 본소의 2차 표준중성자교정실조건에 의하여 이론적으로 예측한 값과 비교하였다. 비교한 결과 실험하여 얻어진 상기의 3가지 계측기에 대한 교정실산란보정인자가 이론적으로 예측한 결과와 최대 약 10% 이내에서 일치하였다.