A Survey on Patient Dose and Exposure Conditions in Simple Radiography of the Abdomen

복부 단순 X-선 촬영조건과 환자 피폭에 관한 조사 연구

  • Kim, Sung-Soo (Dept. of Radiotechnology, Shin Gu College) ;
  • Lee, Sun-Sook (Institute of Radiological Technology, Dong A X-ray Co., LTD.) ;
  • Huh, Joon (Institute of Radiological Technology, Dong A X-ray Co., LTD.)
  • 김성수 (신구전문대학 방사선과) ;
  • 이선숙 (동아엑스선기계 방사선기술연구소) ;
  • 허준 (동아엑스선기계 방사선기술연구소)
  • Published : 1996.12.20

Abstract

We studied exposure techniques and exposure dose for simple abdomen A-P projection for 41 medical facilities that are located in Seoul area. 1. The range of tube voltage used was 60 to 84 kVp, the average tube voltage used was 74 kVp 2. Only 17% of added filter was used. 3. Tube current mostly used was 200 mA, some of them used 400 mA. 4. The grid ratio mostly was used 10 : 1, 54 % of the rare earth screen was used in most facilities. 5. The average skin entrance dose was 4.15 mSv and the dose range was 1.05 mSv to 11.0 mSv.