The Influence of Collimators on SPECT COR Measurements

SPECT 회전중심 측정에 있어서 콜리메타의 영향

  • Lee, Man-Koo (Dept. of Radiotechnology, Wonkwang Public Health College)
  • 이만구 (원광보건전문대학 방사선과)
  • Published : 1996.06.30

Abstract

Misalignment between the electronic and mechanical axes of rotation will result in artifact generation and image degradation during single photon emission computed tomography(SPECT) reconstruction. Acceptance and quality control testing procedures have not emphasized the variability in center of rotation(COR) measurements caused by collimators and the need to verify uniformity across the full collimator field of view (FOV). Variation from the mean COR across the FOV was tested in four different collimators using multiple point source acquisitions. The mean COR was different for each collimator and one of the four had a > 0.5 pixel difference from the mean COR on some area of the FOV, This variation makes this collimator unacceptable for SPECT acquisition. Thus, initial acceptance testing of SPECT collimators should verify a uniform COR across the full FOV and collimators with a variability from the mean COR > 0.5 pixels should be rejected.