Accuracy in target localization in stereotactic radiosurgery using diagnostic machines
정위적 방사선수술시 진단장비를 이용한 종양위치결정의 정확도 평가
Abstract
The accuracy in target localization of CT, MR, and digital angiography were investigated for stereotactic radiosurgery. The images using CT and MR were obtained out of geometrical phantom which was designed to produce exact coordinates of several points within a 0.lmm error range. The slice interval was 3mm and FOV was 35cm for CT and 28cm for MR. These images were transferred to treatment planning computer using TCP/IP in forms of GE format. Measured 3-D coordinates of these images from planning computer were compared to known values by geometrical phantom. Anterior-posterior and lateral films were taken by digital angiography for measurement of spatial accuracy. Target localization errors were 1.2
정위적 방사선수술을 시행하기 위하여 CT, MR, 그리고 digital angiography 의 표적 위치결정의 정확도를 평가하였다. 특정한 지점의 위치가 이미 알려져 있는 (정밀도 : 0.lmm 이내) geometrical phantom 을 CT 혹은 MR로 scan 하였다. Scan 간격은 3mm 이었으며 FOV (Field of View)는 CT의 경우 35cm, MR 의 경우 28cm 이었다. 얻어진 영상은 GE format 의 형태로 TCP/IP를 이용하여 치료계획 컴퓨터로 전달되었다. 각 영상을 컴퓨터로 분석하여 얻은 측정된 좌표값을 이미 알고 있는 값과 비교하였다. Digital angiography 의 경우, 좌표값을 결정하기 위하여 anterior-posterior 방향과 lateral 방향의 film을 얻었으며 이 film은 film scanner를 이용하여 치료계획 컴퓨터로 전달되었다. CT 영상의 위치결정 오차는 1.2