초록
The automated ECG diagnostic systems in hospital have a low P-wave detection capacity in case of some diseases like conduction block. The purpose of this study is to improve the P-wave detection ca- pacity using wavelet transform. The first procedure is to remove baseline drift by subtracting the median filtered signal from the original signal. The second procedure is to cancel ECG's QRS-T complex from median filtered signal to get P-wave candidate. Before we subtracted the templete from QRS-T complex, we estimated the best matching between templete and QRS-T complex to minimize the error. Then, wavelet transform was applied to confirm P-wave. In particular, haiti wavelet was used to magnify P-wave that consisted of low frequency components and to reject high frequency noise of QRS-T complex cancelled signal. Finally, p-wave was discriminated and confirmed by threshold value. By using this method, We can got the around 95.1% P-wave detection. It was compared with contextual information.