Development and Testing of CdZnTe Detector for Pocket Surveymeter

CdZnTe 검출기를 이용한 개인용 Pocket Surveymeter의 제작 및 특성

  • Published : 1996.03.30

Abstract

In this paper, we discussed the fabrication and characterization of bulk type CdZnTe detector for pocket surveymeter. The resistivity of CdZnTe single crystal grown by the High Pressure Bridgman method is in the mid of $10^9$ ohm-cm. The detector structure is Au/CdZnTe/Au and gold electrode is formed by electroless deposition method. Resolutions of 4.8keV and 2.2keV were observed at 22.2keV line of $^{109}Cd$ and 59.6keV line of $^{241}Am$ at room temperature, respectively. We also constructed the small size pocket surveymeter using home made CdZnTe detector. It shows the good linearity over a range from 1mR/hr to 10R/hr with deviation less than 5%. The sensitivity of the surveymeter developed is $2.2{\times}10^3 cps/Rad\;hr^{-1}$ for the 662keV of $^{l37}Cs\;{\gamma}-ray$.

본 논문에서는 bulk형 CdZnTe 감마선 검출기의 제작과 이를 이용한 개인용 선량율계의 제작 및 그 특성에 관하여 기술하였다. 감마선 검출기는 고압 Bridgman법으로 성장된 비저항이 $10^9ohm-cm$이상인 단결정을 사용하였으며 electroless deposition법으로 금전극을 형성시켜 사용하였다. 제작된 CdZnTe 검출기는 $^{109}Cd$의 22.2 keV 감마선과 $^{241}Am$의 59.6 keV 감마선에 대하여 상온에서 각각 4.8keV와 2.2keV의 분해능을 보였다. 또한 이 검출기를 이용하여 개인용 선량율계를 제작하였는데 662keV의 $^{l37}Cs$의 감마선에 대하여 1mR/hr에서 10R/hr의 선율에서 변동율 5%이하의 좋은 직진성을 보였고 온도변화 및 조사선율의 각도분포에 대하여도 좋은 응답 특성을 보였다.