Automatic On-line BOD Measurement System Using A Microbial Membrane Electrode

미생물막 전극을 이용한 BOD 자동화 측정장치

  • Oh, Hyuk (Research Institute for Basic Sciences and Department of Chemistry, Kyung Hee University) ;
  • Kim, Hai Dong (Research Institute for Basic Sciences and Department of Chemistry, Kyung Hee University)
  • 오혁 (경희대학교 기초과학연구소, 문리과대학 화학과) ;
  • 김해동 (경희대학교 기초과학연구소, 문리과대학 화학과)
  • Received : 1995.09.11
  • Published : 1995.09.25

Abstract

Automatic biochemical oxygen demand(BOD) measurement system has heen developed using a microbial membrane electrode, prepared from Bacillus subtilis and polyvinyl alcohol(PVA). The automatic BOD measurement system showed a linear response curve up to BOD 60 ppm using a glucose/glutamic acid standard solution, and all the BOD measurement processes are carried out automatically to calculate BOD whithin 10 min after each sample injection. The response times of the microbial electrode was 5 minutes with a 5 min recovery time between measurements and the relative error of the BOD estimation was within 10%.

Bacillus subtilis와 PVA를 사용한 미생물막 전극을 이용하여 수질시료의 생화학적 산소요구량(BOD)을 측정할 수 있는 BOD 자동화 측정장치를 개발하였다. 본 연구에서 개발한 BOD 자동화 측정장치는 glucose/glutamic acid BOD 표준용액을 사용시 60ppm까지 직선적인 감응을 나타내었으며, 시료주입 후 10분 이내에 자동으로 BOD 측정기능을 수행하였다. Bacillus subtilis와 PVA를 사용한 미생물막 전극의 감응시간과 회복시간은 각각 5분이었으며, BOD 측정 정장치의 자동화에 의하여 10% 오차내에서 재현성 있는 결과를 얻을 수 있었다.