Fundamental Studies of Gas-jet Assisted Radio Frequency Glow Discharge Atomic Absorption Spectrometry

원자흡수법에 의한 Gas-Jet Assisted RF 글로우방전 시스템의 특성 연구

  • Choi, Seong-Kyu (Department of Pharmacy, DongDuck Women's University) ;
  • Kim, Hyo-Jin (Department of Pharmacy, DongDuck Women's University)
  • 최성규 (동덕여자대학교 약학대학) ;
  • 김효진 (동덕여자대학교 약학대학)
  • Received : 1995.01.10
  • Published : 1995.03.25

Abstract

The instrumentation and operating characteristics of a gas-jet assisted radio frequency glow discharge atomic absorption spectrometry(RF-GDAAS) are studied. The roles of discharge power, pressure as well as gas flow rate on the sample loss rate and absorbance are described in this paper. Results show that the optimum discharge conditions depend on pressure and discharge power as well as on the gas flow rate. The gas flow rate plays as an important parameter in RF-GDAAS. Absorbance increases as the gas flow rate increase up to 600ml/min., and the optimum absorbance is obtained at the pressure of 3 mbar. The absorbance increases when the discharge power increases as expected, however, the instability of the system due to the heat might be the trade off factor. The linear calibration curves of Ni and Cr in stainless steel(BAS 461-468) and brass (KSS 1121-1126) are obtained with RF-GDAAS.

원자흡수용 gas-jet assisted RF 글로우 방전시스템을 제작하고 시료손실량과 원자흡광도에 미치는 실험변수들에 관한 연구를 수행하였다. 시료손실량과 흡광도는 방전전력, 압력 및 가스흐름에 큰 영향을 받았다. 원자흡광도와 시료손실량을 가스흐름이 600ml/min로 증가할 때 까지 계속 증가하였으며, 방전 압력은 3mbar에서 최대 흡광도를 나타내었다. 방전전력이 증가할수록 시료손실량과 흡광도는 증가하였으나 시스템의 안정도에 영향을 미치는 관계로 적당한 값의 선택이 필요하였다. 스테인리스 스틸 시료 중의 Ni과 황동시료 중의 Cr의 검량선을 작성한 결과 양호한 직선성이 얻어졌다.

Keywords

Acknowledgement

Supported by : 한국과학재단