직접 디지탈 방사선 촬영시스템과 Ektaspeed 및 Ektaspeed Plus 필름을 이용한 방사선 사진용 디지탈 영상시스템과의 비교 연구

COMPARATIVE STUDY OF DIRECT DIGITAL RADIOGRAPHIC SYSTEM WITH FILM-BASED DIGITAL IMAGING SYSTEM USING EKTASPEED AND EKTASPEED PLUS FILM

  • 도정주 (단국대학교 치과대학 구강악안면방사선학 교실) ;
  • 김은경 (단국대학교 치과대학 구강악안면방사선학 교실)
  • Do Jung-Joo (Dept. of Oral and Maxillofacial Radiology, College of dentistry, DanKook University) ;
  • Kim Eun-Kyung (Dept. of Oral and Maxillofacial Radiology, College of dentistry, DanKook University)
  • 발행 : 1995.06.01

초록

The purpose of this investigation was to compare the direct digital radiographic system with film-based digital imaging system using Ektaspeed and Ektaspeed Plus film with respect to image characteristics and detectability and evaluate the sensor noise with the use of subtraction method. Direct digital radiographic system which used was Sens-A-Ray system(Regam Medical Systems, Sundsvall, Sweden) and film-based digital imaging system was composed of Macintosh II ci computer, high resolution Sony XC-77 CCD camera and intraoral x-ray film(Kodak Ektaspeed film, Kodak Ektaspeed Plus film). Images were taken by using CCD sensor of Sens-A-Ray system, Ektaspeed film and Ektaspeed Plus film with variable exposure time(0.06s, 0.1s, 0.16s, 0.2s, 0.3s, 0.4s, 0.5s, 0.6s, O.8s, LOs), 5 times at each exposure time. And then ektaspeed films and ektaspeed plus films were digitized using CCD camera. Image groups were divided into 3 groups; Sens-A-Ray group(direct digital radiographic system), Ektaspeed group and Ektaspeed Plus group (film-based digital imaging system) They were assessed by the following three aspects; image density, image contrast and detectability and sensor noise of Sens-A-Ray system was also evaluated. The results were as follow : 1. S group showed higher density than E , EP group except at the low exposure time(p<0.01). 2. S group showed higher contrast than E,EP group except at the high exposure time(p<0.01). 3. All groups showed good detectability at the each proper exposure time. Lowest exposure time which shows maximum detectability in S,EP group(0.5s) was lower than that in E group(0.6s). 4. Sensor noise of Sens-A-Ray system generally increased according to exposure time. On the basis of the above results, it was considered that Sens-A-Ray system could show higher speed, higher contrast than Ektaspeed, Ektaspeed Plus film except at too high and low exposure time and the same detectability as the conventional intraoral film.

키워드