Measurement of Traceability Error for Calibration Service Center Using Type S Thermocouples

S형 열전대를 이용한 교정검사기관의 소급성 오차측정

  • Published : 1995.02.28

Abstract

We measured the traceability error for nine high temperature calibration service centers including KRISS through the round-robin test. In this test the type S thermocouple, which used as a calibration standard thermometer commonly, was accommodated as a test thermocouple. Intercomparison data of three institutions were coincident with KRISS's data within ${\pm}0.5^{\circ}C$, which was the calibration uncertainty of the type S thermocouple, but the remaining six institution's data were deviated from the assigned uncertainty level. Deviation of the intercomparison data increased gradually according to the increase of the test temperature. and the maximum difference was so large as about $2.0^{\circ}C$ at the highest test temperature, gold point. In this study we found the traceability error of high temperature calibration service center for a high temperature standard was within $2.0^{\circ}C$.

고온에서 교정용 표준기로 사용되는 S형 열전대를 사용하여, 국내의 9개 고온분야 교정검사기관의 고온측정 능력을 순회교정을 통하여 평가하여 고온의 소급성 오차를 측정하였다. 참여기관 중 3개교정검사기관만이 열전대의 교정불확도인 ${\pm}0.5^{\circ}C$이내에서 한국표준과학연구원의 측정값과 일치하였으며, 나머지 6개 기관의 측정값은 규정된 불확도를 벗어났다. 각 측정값의 차이는 온도가 높아질수록 증가하여 비교측정점 중 가장 높은 온도인 금점에서 가장 큰 차이가 $2^{\circ}C$로 교정 불확도를 크게 벗어났다. 이 연구에서 측정된 국내 고온분야 교정검사기관의 소급성 오차는 $2.0^{\circ}C$ 이내였다.