A Study of the Medical Exposure Dose in Abdomen A-P X-ray Examination

복부(腹部) 단순X선검사시(單純X線檢査時) 피검자(被檢者)의 피폭선량(被曝線量)에 대(對)한 연구(硏究)

  • Kim, Chang-Kyun (Dept. of Radiotechnology, Junior College of Allied Health Science, Korea University)
  • 김창균 (고려대학교 보건전문대학 방사선과)
  • Published : 1994.05.20

Abstract

This study was conducted to find out the medical exposure dose in simple abdomen A-P projection of adults, based on the 87 hospitals located in Seoul. As the results, the following conclusions have been reached; 1. 88.5 % of the surveyed hospitals had the use of $65\;kVp{\sim}79\;kVp(M{\pm}SD:71.45{\pm}4.73\;kVp)$ as tube voltage. 2. 87.35 % of the surveyed hospitals had the use of $50\;mAs{\sim}89\;mAs(M{\pm}SD:64.31{\pm}16.21\;mAs)$ as the amount of current. 3. Shallow doses ranged from 2.00 mSv to 4.99 mSv($M{\pm}SD:3.81{\pm}1.01\;mSv$) in 80.46 % of the surveyed hospitals. 4. Exposure dose was directly depended on the tube voltage or the amount of currents.