Simultaneous Determination of Chromium (III) and Chromium(VI) by High Performance Liquid Chromatography(HPLC)

고성능 액체크로마토그래피(HPLC)를 이용한 3가, 6가 크롬의 동시정량에 관한 연구

  • Roh, Jae Hoon (Institute for Occupational Health, Yonsei University College of Medicine) ;
  • Kim, Chi Nyon (Institute for Occupational Health, Yonsei University College of Medicine) ;
  • Kim, Choon Sung (Institute for Occupational Health, Yonsei University College of Medicine) ;
  • Kim, Kyoo Sang (Institute for Occupational Health, Yonsei University College of Medicine)
  • 노재훈 (연세대학교 의과대학 산업보건연구소) ;
  • 김치년 (연세대학교 의과대학 산업보건연구소) ;
  • 김춘성 (연세대학교 의과대학 산업보건연구소) ;
  • 김규상 (연세대학교 의과대학 산업보건연구소)
  • Published : 1994.11.30

Abstract

Analytic methods for Cr(VI) level in industrial hygienic field were suggested by the National Institute for Occupational Safety and Health(NIOSH method 7600, 7604). There were growing needs for measurement of Cr(III) and Cr(VI) levels simultaneously. Two analytical methods were suggested to determine Cr(III) and Cr(VI) levels simultaneously. The one is method by using reversed phase high peformance liquid chromatography(HPLC) and the other is by using ion exchange HPLC. The purpose of this work was to evaluate the usefulness of these two analytic methods. For the difference of ionic charges of Cr(III)-ethylendiamine tetraacetic acid(EDTA) chelate and $CrO_4{^-2}$, we could detect them simultaneously by ion exchange HPLC. Also, we attempted to determine the levels of Cr(III) and Cr(VI) chelated with sodium diethyldithiocarbamate(NaDDTC) by using reversed phase HPLC. The confirmation of Cr(III) and Cr(VI) were checked by fraction collector and nameless atomic absorption spectrometer. The optimal conditions for the formation of Cr(III)-EDTA chelate were two hours incubation period with pH 5. Cr(III)-EDTA and Cr(VI) in EDTA solution were successfully separated by anion exchange column using $Na_2CO_3/NaOH$ mixture as mobile phase. Peaks of Cr(III)-EDTA and Cr(VI) in EDTA were identified at 5 minutes and 7 minutes of retention time respectively by the ion exchange HPLC. The formation of Cr(III)-NaDDTC and Cr(VI)-NaDDTC chelates were twelve hours incubation period. Cr(III)-NaDDTC and Cr(VI)-NaDDTC chelates were separated by reversed phase column using methanol and water mixture as mobile phase. Peaks of Cr(VI)NaDDTC and Cr(III)-NaDDTC chelates were identified at 13 minutes and 26 minutes of retention time respectively by the reversed phase HPLC. Due to reduction of Cr(VI) to Cr(III), it seems to be not suitable for simultaneous determination of Cr(III)-NaDDTC and Cr(VI)-NaDDTC chelates by reversed phase HPLS. Simultaneos determination of Cr(III) and Cr(VI) by ion exchange HPLC was more accurate and simple method.

Keywords

Acknowledgement

Supported by : 대한산업보건협회, 산업보건연구소