The evaluation of absorbed dose of rectum for the patient of intracavitary treatment $^{60}$ Co by MgB$_4$O$_{7}$ (Mn) TLD

MgB$_4$O$_{7}$ (Mn) TLD에 의한 $^{60}$ Co 강내 치료 환자의 직장의 흡수선량 평가

  • 유명진 (고신대학교 의과대학 의예과) ;
  • 신병철 (고신대학교 의과대학 의예과, 치료방사선과학교실) ;
  • 문창우 (고신대학교 의과대학 의예과, 치료방사선과학교실) ;
  • 정태식 (고신대학교 의과대학 의예과, 치료방사선과학교실) ;
  • 염하용 (고신대학교 의과대학 의예과, 치료방사선과학교실) ;
  • 도시홍 (수산대학교 물리학과)
  • Published : 1994.12.01

Abstract

By measuring with TLD, absorbed dose of rectum will be determined more accurately than calculated value. MgB$_4$O$\sub$7/(Mn) TLD phosphor was prepared and its characteristic was studied. To measure the absorbed dose of rectum with TLD, 33 patients who had intracavitary treatment with $\^$60/Co from April 1994 to September 1994 at Kosin Medical center. There were 10-59% differecne in most cases between the calculation and the measurement for the absorbed dose of rectum. The difference was less the case with tandem and ovoid than the case with only ovoid.

자궁강내 치료시 직장의 흡수선량을 TLD로 직접 측정함으로써 컴퓨터에 의한 계산값과 비교하여 보다 더 정확한 흡수선량을 결정하고자 한다. 마그네슘 보레이트에 Mn을 활성제로 첨가한 MgB$_4$O$_{7}$ (Mn) TLD를 제작하여 기본적인 특성조사를 하였다. 1994년 4월부터 9윌까지 고신의료원 치료방사선과에서 $^{60}$Co 강내치료를 받은 환자 33명을 대상으로 직장의 흡수선량 측정을 하였다. 직장의 흡수선량을 컴퓨터로 계산한 값과 TLD로 측정한 값을 비교한 결과 대부분이 10-59%이내의 차이를 보였다. Tandem과 Ovoid를 같이 사용한 환자보다 Ovoid만 사용한 환자에게 있어서 차이가 적게 나는 것으로 나타났다.