적외선 체열촬영시스템을 위한 광학계의 구성

Realization of Optic Systems for the Infrared Thermography

  • 이수열 (건국대학교 의과대학 의학공학과) ;
  • 우응제 (건국대학교 의과대학 의학공학과) ;
  • 조민형 (수원대학교 공과대학 전기공학과)
  • 발행 : 1994.03.01

초록

In the conventional infrared imaging system, complex infrared lens systems are usually used for directing collimated narrow infrared beams into the high speed 2-dimensional optic scanner. In this paper, a simple reflective infrared optic system with a 2-dimensionaloptic scanner is proposed for the realization of medical infrared thermography system. It has been experimentally proven that the infrared thermography system composed ofthe proposed optic system has the temperature resolution of $0.1{\circ}C$ under the spatial resolution of 1mrad, the image matrix size of $256{\times}240$, and the imaging time of 4 seconds.

키워드

참고문헌

  1. Hand Clinics v.7 no.1 Thermography of the hand and wrist : Practical applications Frieda Feldman
  2. J. of Neurosurg. v.69 Quanification of thermal asymmetry Sumio Uematsu(et al.)
  3. 대한신경외과학회지 v.19 요추간판 탈출증 환자에서의 컴퓨터 적외선 전신체열 촬영의 의의 김영수;조영은;오성훈
  4. Thermological Methods Joachim-Michael Engel(ed.)
  5. Passive Electro-Optical Systems Stephen B. Campana(ed.)
  6. Analysis, and Testing Electro-Optical System Design Michael C. Dudzik(ed.)
  7. Modern Optical Engineering Warren J. Smith
  8. Thermological Methods Joachim-Michael Engel(ed.)
  9. Electro-Optical Components William D. Rogatto(ed.)
  10. The Infrared Handbook(3rd ed.) George J. Zissis