The Compensation of the Energy Dependence of TLD

열형광선량계(熱螢光線量計)의 에너지의존성(依存性) 보상(補償)

  • Park, Myeong-Hwan (Dept. of Radiotechnology, Taegu Health Junior College) ;
  • Lee, Joon-Il (Dept. of Radiotechnology, Taegu Health Junior College) ;
  • Kwon, Duk-Moon (Dept, of Diagnostic Radiology, Yeungnam University Hospital)
  • 박명환 (대구보건전문대학 방사선과) ;
  • 이준일 (대구보건전문대학 방사선과) ;
  • 권덕문 (영남대학교병원 방사선과)
  • Published : 1993.12.30

Abstract

The $CaSO_4$ : Tm-PTFE TLDs have been prepared and their energy dependences of TL intensity to X-rays have been investigated. The dose dependence of the prepared $CaSO_4$ : Tm-PTFE TLDs is linear within the range of $100{\mu}Gy-10Gy$ for X-rays and gamma rays. The spectral peaks of TL emission spectrum are at about 350nm and 475nm. The fading rate of the main peak has been found to be about 2% for 30days when $CaSO_4$ : Tm-PTFE TLDs stored in the dark room at room temperature. The energy dependence of $CaSO_4$ : Tm-PTFE TLD has been compensated by combining the TL responses of one bare TLD and five TLDs filtered with 1.0mm Al, 0.2mm Cu, 0.5mm Cu, 1.5mm Cu, 2.0mm Pb respectively. The determined correction coefficients for filter combination are $a_0=0.078,\;a_1=-0.009,\;a_2=-0.235,\;a_3=0.588,\;a_4=0.568\;and\;a_5=0.180$ respectively. From the results of these studies, the prepared TLD badge of six dosimeter combination may be useful as a radiation dosimeter for personal and environmental monitoring.

Keywords