진단용(診斷用) X선장치(線裝置)의 관전압(管電壓) 백분율(百分率) 평균오차(平均誤差)에 관한 조사(調査) 연구(硏究)

A Study on PAE of Tube Voltage of Diagnostic X-ray Units -At the area of Sungnam, Inchon, Seoul-

  • 임한영 (인산전문대학 방사선과) ;
  • 오현주 (신구전문대학 방사선과) ;
  • 김영일 (신구전문대학 방사선과)
  • Lim, Han-Young (Dept. of Radiotechnology, In San Junior College) ;
  • Oh, Hyun-Joo (Dept. of Radiotechnology, Sin Gu Junior College) ;
  • Kim, Young-Il (Dept. of Radiotechnology, Sin Gu Junior College)
  • 발행 : 1993.12.30

초록

In this paper, the PAE of tube voltage of 93 diagnostic X-ray units those are using at the 19 general hospitals and 9 local clinics in the area of Sungnam, Inchun, and Seoul have been surveyed. The results are summarized as following. 1. PAE of tube Voltage 70, 100KV at the 56 (60%) X-ray units were ${\pm}7%$ below, at the 8 units (9%), were ${\pm}7%$ above, and those 29 X-ray units (31%) were ${\pm}7%$ below or above with the 70, 100KV tube Voltage in a unit. 2. PAE ${\pm}7%$ below were 68 X-ray units, at the low tube voltage (70KV), and 76 X-ray units at the high tube Voltage (100KV). 3. For six year-use abofe, used X-ray units were PAE ${\pm}7%$ above, and so, the longer used years the more X-ray units became declined in those performance. 4. PAE of tube Voltage were better efficiency at the large capacity 30 X-ray units than small capacity 10. 5. PAE ${\pm}7%$ abovein were only 8 units among 84 diagnostic X-ray units in the general hospitals.