Study on the Plasma Oscillation for Gas Chromatographic detector

플라즈마진동을 이용한 기체크로마토그래프의 검출기에 대한 연구

  • Received : 1993.12.02
  • Published : 1993.10.25

Abstract

Plasma oscillation has been observed during the development of the glow discharge as detector for gas chromatography. The variation of oscillation frequency shows the better stability and detection limits than the changes in the dischange curent. To investigate the range of useful operating conditions and to gain insight into the mechanism, the effect of experimental parameters on plasma oscillation have been studied. This study includes the variation of discharge current, pressure and discharge gap. Frequency ranges of 10KHz to 10MHz have been observed with the various shapes of oscillation. Two kinds of mode for oscillation are observed with the variation of electrode gap at low pressure and low voltage.

플라즈마 진동 현상은 글로우방전을 이용한 기체크로마토그래프 검출기의 개발 과정에서 측정되었으며 전류의 변화에 비하여 매우 안정하며 낮은 검출한계로 새로운 검출기로의 개발 가능성이 높다. 플라즈마 진동의 최적 조건과 메카니즘을 규명하기 위하여 몇 가지 실험조건을 변화시키며 관찰하였다. 즉, 방전전압(전류), 압력, 그리고 방전전극 간격을 변화시킨 결과 다양한 형태의 진동이 10KHz에서 10MHz까지의 주파수 범위에서 관찰되었다. 특히 방전전극 간격에 따라 낮은 압력과 전압의 조건에서 두 가지 형태의 진동이 나타났다. 한 가지 형태는 positive column 영역에서, 또 다른 하나는 negative glow 영역에서 측정되었다. Positive column 영역보다 negative glow 영역에서 발생된 플라즈마 진동이 훨씬 더 높은 감도 변화와 더 큰 주파수를 보였다.