Design and Diafiltration Performance of Modified Cellulose and Regenerated Cellulose Hollow Fiber Membrane Hemodiafilter

개질 셀룰로오즈 및 재생 셀룰로오즈 중공사막 혈액투석여과기의 설계 및 투석여과성능

  • 김재진 (한국과학기술연구원 분리막연구실) ;
  • 박진용 (한국과학기술연구원 분리막연구실, 녹십자의료공업(주) 중앙연구소)
  • Published : 1993.09.01

Abstract

Modules of hemodiafilter were manufactured by using MC-Hp200 and RC-HP4DOA hoi low fiber membranes of Enka Co. in a unique design. The performances of she hemodiafillers were evaluated by measuring the molecular weight cut-off, ultrafiltration rate, clearance, and pressure drop across the hemodiafiter. As a whole, the performances of the RC-HP400A module were superior to those of the MC-Hp200 module. The modules prepared in thls study showed the satisfactory performances for hemodiafiltration.

Keywords

References

  1. Artif. Organs v.5 R.Schmidt;P.Ivanovich;H.Klinkmann
  2. Artif. Organs v.2 H.W.Leber;V.Wizemann;G.Goubeaud;P.Rawer;G.Schutterle
  3. J. Dialysis v.3 W.Kohnel;K.B.G.Sprenger;B.Spohn;H.E.Franz
  4. Blood Purif. v.2 J.A.Van Geelen;W.Carpay;J.A.Flendrig
  5. 의공학회지 v.14 폴리설폰 중공사막 혈액여과기의 설계 및 여과성능 김재진;박진용;서상봉;김성수;김은영;강성종;연인석;고영환
  6. 한국과학기술연구원 연구보고서 Hemodiafilter의 개발 김재진(외)