Clinical Application of Inverter type X-ray Generator

Inverter 방식(方式) X선장치(X線裝置)의 임상응용(臨床應用)

  • Lee, Sun-Sook (Dept. of Radiotechnology, Junior College of Allied Health Sciences, Korea University) ;
  • Huh, Joon (Dept. of Radiotechnology, Junior College of Allied Health Sciences, Korea University) ;
  • Lee, Jea-Won (Dept. of Radiology, Paik Hospital, In Je University) ;
  • Kim, Jung-Min (Dept. of Radiotechnology, Ji San Junior College)
  • 이선숙 (고려대학교 보건전문대학 방사선과) ;
  • 허준 (고려대학교 보건전문대학 방사선과) ;
  • 이재원 (인제대학교부속 백병원 방사선과) ;
  • 김정민 (지산간호보건전문대학 방사선과)
  • Published : 1992.05.30

Abstract

Inverter type X-ray apparatus has been introduced and used several hospitals. Principle of inverter type X-ray generators are such as to convert the frequency of commercial power supply to high frequency and to control the high voltages for X-ray tube. Inverter generators are now on the way for futher development to elliminate single phase generators and three phase generators. We compared inverter type X-ray apparatus with conventional single phase 2 peak and three phase 12 peak, apparatus in the following aspects X-ray out put to tube voltage, linearity of X-ray out put to mA, HVL according to mA contrast to kV.

Keywords