A PHOTOELASTIC STUDY OF STRESS DISTRIBUTIONS AROUND SURROUNDING BONE TISSUES OF IMPLANTS DEPENDING UPON THE LENGTH OF IMPLANT FIXTURES SUPPORTING FIXED BRIDGES

고정성 가공의치를 지지하는 임프란트의 길이에 따른 임플란트 주위 골조직에서의 응력분산에 관한 광탄성 연구

  • Bae Hyong-Su (Department of Prosthodontics, School of Dentistry, Kyungpook National University) ;
  • Cho Sung-Am (Department of Prosthodontics, School of Dentistry, Kyungpook National University)
  • 배형수 (경북대학교 치과대학 보철학교실) ;
  • 조성암 (경북대학교 치과대학 보철학교실)
  • Published : 1992.12.01

Abstract

The Present study was performed to evaluate the stress distribution patterns of bone tissues around two Branemark implant fixtures with varying lengths. Six models were used with the combination of 10mm, 7mm, 5mm Branemark implant fixture analogs. The load distribution characteristics of surrounding bone tissues were observed under 64kg vertical static loading on central part of fixed bridge using 2 dimensional photoelastic technique. Through observing the pattern of stress distribution around the implant fixtures, I got the results as follow : 1. Stress distribution pattern of model I(10mm, 10mm) was best among those of six models. 2. When two fixtures were in equal length, stress was increased in proportion to the reduction of implant surface area. 3. When two fixtures were in unequal length, stress was concentrated on the neck portion of the shorter one.

저자는 두개의 브로네마트 임플란트를 지대치로 하는 가공의치를 제작할 때 임플란트의 길이에 따른 임플란트 주위 조직의 응력분산양상을 평가하기 위하여 임플란트의 길이가 각각 10mm, 7mm, 5mm인 여섯가지의 실험모형을 제작하고 가공의치 중아부에 64kg의 수직하중을 가하여 2차원적 광탄성응력분석 실험을 행하여 비교 분석 해본 바, 다음과 같은 결론을 얻었다. 1. 여섯가지 실험모형 중 응력분산 양상은 실험모형 I (10mm, 10mm)일 경우가 가장 좋았다. 2. 근심과 원심에 같은 길이의 임플란트를 매식하였을 경우, 응력은 대체로 임플란트 표면적의 감소에 비례하여 증가하였다. 3. 근심과 원심에 다른 길이의 임플란트를 매식하였을 경우, 짧은 쪽 임플란트의 경부에 응력이 집중되었다.

Keywords