Development of a Bipolar Pulse Conductometric Detector for Environmental Monitoring and Electrochemical Experiments

환경수질오염 검증과 전기화학실험을 위한 양극펄스형 전기전도성검출기의 개발

  • Kim, Haidong (Advanced Analysis Center, Korea Institute of Science and Technology) ;
  • Kim, Sun Tae (Advanced Analysis Center, Korea Institute of Science and Technology) ;
  • Kim, Taek-Jae (Advanced Analysis Center, Korea Institute of Science and Technology) ;
  • Kim, Young-Sang (Department of Chemistry, Korea University)
  • 김해동 (한국과학기술연구원 특성분석센터) ;
  • 김선태 (한국과학기술연구원 특성분석센터) ;
  • 김택제 (한국과학기술연구원 특성분석센터) ;
  • 김영상 (고려대학교 화학과)
  • Received : 1992.03.11
  • Published : 1992.03.25

Abstract

A bipolar pulse conductometric detector has been developed for continuous on-line monitroring of the flowing sample streams. The conductometric detector was fast and easy to use due to the free of interferences by serial or parallel capacitances. The performance of the bipolar pulse conductometric detector was tested using dummy cells with $10^1{\sim}10^8$ ohm resistance range and found a linear response with a 0.01% relative standard deviation. The possible applications of the bipolar pulse conductometric detector was discussed.

연속적으로 흐르는 시료의 검출을 위한 양극펄스형 전기전도성 검출기를 개발하였다. 개발한 전기전도성 검출기는 동작이 매우 빠를 뿐더러 직렬 혹은 병렬축전 방해효과가 없으므로 사용이 쉽다. 개발한 양극펄스형 전기전도성 검출기의 성능을 $10^1{\sim}10^8$저항을 가진 가상 셀들을 사용하여 시험한 결과 0.01% 상대편차를 보이는 직선성 응답을 얻을 수 있었다. 개발한 전기 전도성 검출기의 장단점과 여러 가능한 응용들에 대해서도 검토하였다.

Keywords