The Polymer Membrane Electrode by Surfactants for Measuring Continuously Thiocyanate Ion in Wastewater

폐수중 티오시안산이온을 측정하기 위한 계면활성제를 이용한 고분자 막전극

  • 최종석 (명지대학교 화학공학과) ;
  • 안형환 (명지대학교 화학공학과) ;
  • 강안수 (명지대학교 화학공학과) ;
  • 우인성 (인천대학교 산업안전공학과) ;
  • 황명환 (인천대학교 산업안전공학과)
  • Published : 1991.12.01

Abstract

Ion-selective electrode responsive to the thiocynate ion prepared by using the quaternary ammonium salts as a active material and PVC as a membrane matrix. The effect of chemical structure and composition of active material, and the membrane thickness on the linear response. the detection limit, and Nernstian slope of the electrode studied. Under the above optimum conditions of membrane, the effect of pH and the selectivity coefficients to various interfering anions were compared and investigated. It was concluded that the functions of thiocynate ion-selective electrode(ISE) were closely related to the chemical structure of the quaternary ammonium salts. The linear response, and the detection limit of the electrode potential increased with the increase of the carbon chain length of the alkyl group in the quaternary ammonium salts in the ascending order of Aliquat 336T, TOAT, TDAT, and TDDAT. The optimum membrane thickness was 0.3mm. The electrode characteristics was better with the decrease of the concentration of active material, and the best concentration was 3 weight percent. The membrane potential was independent of the pH variation in the region from pH 2 to 12. The order of the selectivity coefficients is as follows:Cl $O_4$$^{[-10]}$$I^{[-10]}$ >N $O_3$$^{[-10]}$ >B $r^{[-10]}$$F^{[-10]}$ >C $l^{[-10]}$ >O $A_{c}$ $^{[-10]}$ 〓S $O_4$$^{2-}$.

Keywords