The Calibration of $^{90}$ Sr Ophthalmic Applicator by Measuring Electron Current

전류 측정 방식에 의한 안과용 $^{90}$ Sr 선원의 교정

  • 이병용 (아산재단 서울중앙병원, 울산대학교 의과대학 치료방사선과학교실) ;
  • 신동오 (경희대학교 부속병원, 경희대학교 의과대학 치료방사선학교실) ;
  • 김현자 (부산침례병원 치료방사선과) ;
  • 홍성언 (경희대학교 부속병원, 경희대학교 의과대학 치료방사선학교실) ;
  • 최은경 (아산재단 서울중앙병원, 울산대학교 의과대학 치료방사선과학교실) ;
  • 장혜숙 (아산재단 서울중앙병원, 울산대학교 의과대학 치료방사선과학교실)
  • Published : 1991.10.01

Abstract

We have designed and applied the calibrationmethod of $\^$90/Sr Ophthalmic Applicaton by measuring the electron currents. We considered the number of electrons which is emitted from the source, the area of the source, and the electron stopping power in the water, and those data were used for calculation. Film was used for evaluating the accurate source area. Average electron stopping power was obtained by analyzing ${\beta}$-ray energy spectrum. We compared between the result from our method and that from the TLD measurements. The calibration result from our method shows 63.3 ${\pm}$5.1 cGy/sec, while 50.7${\pm}$7.3 cGy/sec from TLD measurement. But the supplier's specification tells 46.89.4cGy/sec.

안과용 $^{90}$ Sr 선원에서 방출하는 $\beta$선의 갯수를 직접 측정함으로써 (전류 측정)표면 선량을 계산하는 방법을 고안하여 실제 교정에 응용하였다. 안과용 $^{90}$ Sr선원에서 방출하는 전체 $\beta$선의 갯수 및 선원의 면적, 전자의 물속에서의 저지능 등을 고려하였다. 선원의 정확한 면적 측정은 필림을 이용하였으며, 평균 저지능은 상기 선원에서 방출되는 $\beta$선의 에너지 스펙트럼을 분석함으로써 얻을 수 있었다. 이 방법과 $^{90}$ Sr으로 교정한 TLD칩을 이용한 교정 방법과의 결과를 비교하였다. 전류 측정 방식에 의한 결과는 63.5$\pm$5.1cGy/sec이었고, TLD칩에 의한 측정 결과는 50.7$\pm$7.3cGy/sec로 오차 범위내에서 일치하였다. 그러나 제작사 보증서에 의하면 46.8$\pm$9.4cGy/sec로 측정값에 비하여 36%의 차이를 보여 주었다.

Keywords