Element Correction Method of Thermoluminescent Dosimeters

개인 피폭선량계 소자 보정법

  • 송명재 (한국전력공사 기술연구원)
  • Published : 1991.04.01

Abstract

Generally, it is an accurate radiation measurement technique for processors fo thermoluminescent dosimenters(TLDs) to characterize each element they use by producing element correction factors(FCFs). TLDs are classified into three groups such as reference. control, and field TLDs. Reference TLDs are used only for the production of ECFs for the control and field TLDs. They are kept locked in a safe place except when it is necessary to use a subset of them to produce ECFs for the control and field TLDs. The ECF of a given element is a measure of the response of the element relative to the mean response of an arbitrarily selected group of reference elements. As TLDs are used in the field, their relative responses to radiation might be decreased due to muliple readings and physical abuse. Therefore, the producditon of ECFs are performed initially and periodically during the field use. This element correction method provides an excellent tool to examine new TLDs and to monitor the reliability of old TLDs. This paper discuss the 10 step procedures developed to produce and examine ECFs.

방사선 피폭량을 정확히 측정하는 방법의 하나로 열형광선량계의 각 소자별 보정인자를 구하여 사용하는 방법이 있다. 열형광선량계를 기준 선량계, 제어선량계 및 현장 선량계의 3그룹으로 분류하여, 기준 선량계는 제어 및 현장 선량계의 소자보정인자 산출시에만 사용하며, 소자보정인자 산출을 위해 일부분씩 사용할 때를 제외하고는 안전한 장소에 보관한다. 소자보정인자는 기준 선량계의 반응도 평균치에 대한 각 소자의 반응도 비로서 정의된다. 선량계는 판독 횟수가 증가하거나 손상으로 인해 반응도가 감소하기 때문에 최초 구입할 때 또는 사용중에 소자보정 인자를 주기적으로 재산출하여야 한다. 이 소자보정방법은 새로운 선량계의 판독 오차를 줄이고 사용중인 선량계의 신뢰도를 향상시킬 수 있는 우수한 방법이다. 본 논문에서는 소자보정인자를 산출하기 위한 10가지 단계별 방법을 자세히 소개하였다.