국산 막형 산화기 (OXYREX)의 체외순환 임상 성적 - 개심수술 40례 적용 -

Clinical Application of OXYREX Hollow Fiber Membrane Oxygenator

  • 발행 : 1990.12.01

초록

The OXYREX hollow fiber membrane oxygenator developed by joint work of KIST and Green Cross Medical company has been evaluated by experimental investigation and clinical application, In this oxygenator gas exchanges occur through small pores of 0.1pm size which are distributed on 70% of surface of polypropylene hollow fiber. The Oxyrex membrane oxygenator consists of 36 thousand hollow fibers and it has 3.3m2 of gas exchange surface. The Oxyrex membrane oxygenator has unique blood flow path: blood enters the oxygenator passes between the hollow fibers and exits through outlet ports, that provides low transmembrane pressure drop. In the animal experiment and in vitro investigations of Oxyrex oxygenator, it showed low transmembrane pressure difference, effective heat exchanger performance, stable gas transfer function and less blood trauma. The Oxyrex oxygenator been used from March, 1990, to October, 1990, in 40 patients undergoing open heart operations. In the clinical applications of Oxyrex, adequate oxygenation[PaO2, 283$\pm$70mmHg] and carbon dioxide removal[PaCO2, 27\ulcorner6mmHg]were maintained under the condition of FiO2: below 0.6, Hct; 25%, perfusion flow; 2.4 L/min, gas flow: 2.1 L/min. During maximum 365 minutes of cardiopulmonary bypass[CPB] time period, the Oxyrex oxygenator maintained stable condition of PaO2, PaCO2 respectively and it also kept low plasma hemoglobin level. The complement proteins C3 and CH50 were not significantly changed pre to post CPB. There were no complications related to the oxygenator during and after the CPB.