Frequency Responses of Ambulatory ECG Systems and Their Correction by a Compensation Circuit

  • Iwao, K. (Fujita-Gakuen Health University) ;
  • Yokoi, A. (Fujita-Gakuen Health University) ;
  • Suzuki, S. (Fujita-Gakuen Health University) ;
  • Goto, T. (Fujita-Gakuen Health University) ;
  • Doniwa, K. (Fujita-Gakuen Health University) ;
  • Okajima, M. (Fujita-Gakuen Health University)
  • Published : 1990.06.01

Abstract

Frequency responses of ambulatory electrocardiogram systems were measured. Phase shift was assessed by our improved Wagner's method. The characteristics of the systems were found much worse than that of ordinary ECG equipment both in gain and phase responses. The decay of -3dB in amplitude observed in 0.2Hz and the lead of 45 dog in phase was observed in 0.6Hz. In order to investigate which of these 2 responses play major role in generating false S -T deviation, mathe- matical filters were Composed in a computer and actual ECGs were fed. The false S-T deviation were found to be caused mostly by enormous lead in phase, and then, the compensation circuit to diminish the lead in phase was inserted in the commercial systemes. The compensated systems showed no false deviation in S-T segment.

Keywords