The calibration of 1 inch standard condenser microphone is done by the reciprocity calibration method. There are two kinds of reciprocity calibration, free-field calibration and pressure calibration. The pressure sensitivities of the three 1 inch condenser microphones are determined by pressure calibration. The accuracy of the pressure sensitivity of the microphone depends on the accuracies of the voltage and dimension measurements as well as the various corrections for the coupler. If the individual accuracies for the measurements and corrections are achieved, it is estimated that the over-all accuracy is approximately 0.05dB at low and middle frequencies decreasing to about 0.ldB over 10kHz.

표준 컨덴서 마이크로폰의 정밀 교정은 가역원리를 이용한 가역교정에 의해 이루어진다. 본 연구에서는 가역교정 중에서도 결합기를 이용하는 음압교정(pressure calibration)을 수행하여 세 개의 1인치 컨덴서 마이크로폰의 음압감도를 결정하였다. 마이크로본의 음압감도를 정밀하게 결정하기 위해서는 전압이나 결합기의 치수를 정밀하게 측정하여야 하며 결합기의 구조상 발생하는 제반 현상들에 대한 보정을 정확하게 해 주어야 한다. 이와같은 사항들을 정화하게 고려할 경우 가역교정에 의한 교정정밀도는 저주파와 중간 주파수 영역에서 0.05dB이며 10kHz 이상에서는 0.ldB가 된다.