A Study on the X-Ray Imaging using Dusl Energy Method

이중에너지 방법을 이용한 X선 영상법에 관한 연구

  • Published : 1988.12.01

Abstract

The dual-energy technique win used to separate the bone-only and tissue-only images from the conventional chest images. The equivalent thickness of the basic materials are estimated from low and high energy images of a given complex materials using the attenuation coefficient of ma serial componens. We showed that the image quality of dual-energy imaging method can be influenced by the ponlinearity and noise components of system and spectrum distributions The quantitative analysis of Calcium component was performed by dual-energy technique and it is shown that the concentration of the Calcium could be accurately estimated within 5% error range.

Keywords