원자력병원 중성자선치료기의 물리적특성

Dosimetric Characteristics of the KCCH Neutron Therapy Facility

  • 류성렬 (한국에너지연구소 원자력병원 치료방사선과) ;
  • 노성우 (한국에너지연구소 원자력병원 치료방사선과) ;
  • 정현우 (한국에너지연구소 원자력병원 치료방사선과) ;
  • 조철구 (한국에너지연구소 원자력병원 치료방사선과) ;
  • 고경환 (한국에너지연구소 원자력병원 치료방사선과) ;
  • 박주식 (싸이클로트론 연구실) ;
  • 줄리 인마 (미국워싱턴 대학 치료방사선과)
  • Yoo Seong Yul (Department of Therapeutic Radiology, Korea Cancer Center Hospital) ;
  • Noh Sung Woo (Department of Therapeutic Radiology, Korea Cancer Center Hospital) ;
  • Chung Hyun Woo (Department of Therapeutic Radiology, Korea Cancer Center Hospital) ;
  • Cho Chul Koo (Department of Therapeutic Radiology, Korea Cancer Center Hospital) ;
  • Koh Kyoung Hwan (Department of Therapeutic Radiology, Korea Cancer Center Hospital) ;
  • Bak Joo Shik (Laboratory of Cyclotron Korea Cancer Center Hospital) ;
  • Eenmaa Juri (Department of Radiation Oncology, University of Washington)
  • 발행 : 1988.06.01

초록

한국에너지연구소 원자력병원 싸이클로트론에 의해 생산되는 중성자를 임상에 적용시키기 위해, 이의 물리적 특성을 알기 위하여 방사선선량 측정실험을 시행하였다. 여기서 얻은 결과를 외국의 다른 치료기관에서 얻은 데이타와 비교 분석하였다. 중심축 선상의 심부선량백분율, build-up곡선, open과 쐐기등선량 곡선의 값이 4MV와 6MV X-ray값의 중간에 위치하였다. 최대선량의 build-up은 피부아래 1.35cm에 위치했으며 입사 선량은 약 $40\%$였다. 출력인자는 $6\times6cm$의 조사야에서 0.894, $30\times30cm$의 조사야에서는 1.187이었다. 중성자선의 X-ray오염도는 $10\times10cm$ 조사야에서 심부 2cm에서 $4.9\%$였다.

For the physical characterization of neutron beam, dosimetric measurements had been performed to obtain physical data of KCCH cyclotron-produced neutrons for clinical use. The results are presented and compared with the data of other institutions from the literatures. The central axis percent depth dose, build-up curves and open and wedge isodose curve values are intermediate between that of a 4 and 6 MV X-rays. The build-up level of maximum dose was at 1.35cm and entrance dose was approximately $40\%$. Flatness of the beam was $9\%$ at Dmax and less $than{\pm}3\%$ at the depth of $80\%$ isodose line. Penumbra begond the $20\%$ line is wider than corresponding photon beam. The output factors ranged 0.894 for $6\times6cm$ field to 1.187 for $30\times30cm$ field. Gamma contamination of neutron beam was $4.9\%$ at 2 cm depth in $10\times10cm$ field.

키워드