Relationship between the Measurement Values of the Digital Dust Indicator and Personal Dust Sampler

Digital 분진계와 개인용 분진 포집기의 측정값의 관계

  • Yoon, Young-No (The Institute of Pneumoconiosis, Korea Labour Welfare Corporation) ;
  • Lee, Jeong-Joo (The Institute of Pneumoconiosis, Korea Labour Welfare Corporation) ;
  • Chung, Ho-Keun (The Institute of Pneumoconiosis, Korea Labour Welfare Corporation) ;
  • Choi, Ho-Chun (The Institute of Pneumoconiosis, Korea Labour Welfare Corporation)
  • 윤영노 (근로복지공사 진폐연구소) ;
  • 이정주 (근로복지공사 진폐연구소) ;
  • 정호근 (근로복지공사 진폐연구소) ;
  • 최호춘 (근로복지공사 진폐연구소)
  • Published : 1987.11.01

Abstract

The digital dust indicator (Sibata P-5), one of the direct·reading instruments was evaluated for the respirable dust in the underground coal mine environments. As a reference, respirable dust was determined using three cyclones and/or impactors. All the tests were performed on aerosol in twenty underground coal mines. The coefficients of mass-relative concentration were $0.067{\pm}0.054$ (Mean$\pm$Standard deviation) (range: 0.006-0.172). The relationship between relative concentration and temperature was not significant statistically. Also, the relationship of relative concentration and relative humidity was not significant. Mass concentration and relative concentration were $5.31{\pm}5.22mg/m^3$ and $162{\pm}163$ CPM ($Mean{\pm}Standard$ deviation) respectively. The range of mass concentration was $1.22-22.69mg/m^3$; relative concentration 16-628 CPM. The relationship of mass concentration and relative concentration was not significant in these ranges.

Keywords