A Study on Equivalent Square Field in High Energy Photon Beam

고에너지 Photon Beam에 있어서 Equivalent Square Field에 관한 고찰

  • Kim, Sung-Kyu (Department of Therapeutic Radiology, College of Medicine, Yeungnam University) ;
  • Shin, Sei-One (Department of Therapeutic Radiology, College of Medicine, Yeungnam University) ;
  • Kim, Myung-Se (Department of Therapeutic Radiology, College of Medicine, Yeungnam University)
  • 김성규 (영남대학교 의과대학 치료방사선과학교실) ;
  • 신세원 (영남대학교 의과대학 치료방사선과학교실) ;
  • 김명세 (영남대학교 의과대학 치료방사선과학교실)
  • Published : 1987.12.30

Abstract

An analytic expression for equivalent square fields and a table are derived using the empirical representations obtained from the measured field size dependence of output for square field in a phantom. The expression is applicable to radiation fields that consistent with the following two assumptions Imposed on these representations. 1) It is a linear function of the logarithm of the field area, and 2) It is approximately the same function for both square and circular field of similar areas. In this paper, the derived tables and BJR table were consulted.

본 병원 18MeV 선형가속기에 대한 equivalent square field를 측정하여 다음과 같은 결과를 얻었다. 1. $10{\times}10cm$ 이하의 작은 면적에서 매우 가늘고 긴 조사에서는 실제 측정치와 약 0.5%의 차이를 나타내고 있다. 2. $10{\times}10cm$보다 크며 $15{\times}15cm$보다 작은 중간 면적에서는 실제 측정치와 약 0.64 %의 차이를 나타내고 있다. 3. $15{\times}15cm$보다 큰 면적에서는 실제 측정치와 약 0.24%의 차이를 나타내고 있다. 위와 같은 측정결과로 (6)식에 의해서 구해진 Table 1의 equivalent square field는 매우 바람직함을 알 수 있으며, 특히 매우 가늘고 긴 작은 면적에 대해서는 BJR table보다 오차를 훨씬 줄이고 있어서 식도암 등을 조사할 시에는 매우 유용함을 알 수 있다. 더 나아가 전 영역을 포함하는 더욱 더 근시적인 equivalent square field method가 연구되어야 할 것이다.