대한의용생체공학회:의공학회지 (Journal of Biomedical Engineering Research)
- 제8권2호
- /
- Pages.161-170
- /
- 1987
- /
- 1229-0807(pISSN)
- /
- 2288-9396(eISSN)
Development of the Rolling-cylinder Type Motor-driven Total Artificial Heart System
- Min, Byoung-G. (Dept. of Biomedical Eng., Seoul National University) ;
- Kim, Hee-C. (Dept. of Biomedical Eng., Seoul National University) ;
- Cheon, Gill-J. (Dept. of Biomedical Eng., Seoul National University)
- 발행 : 1987.12.01
초록
A new type of motor-driven total artificial heart system with a rolloing-cylinder mechanism has been developed. The prototype system consists of a brushless DC motor inside of a rolling-cylinder, two arc shaped pusher-plate s, and two ventricles of smooth, seamless polyurethane sacs. The motor-driven pump has the advantages of being portable and quiet compared to the present air-driven pump. It can also be controlled more accurately. This rolling-cylinder type electromechanical pump has several structural advantages including small size and weight, as compared to other research groups' motor-driven pumps. The results of mock circulation tests confirm sufficient pump output capacity(cardiac output . 9 L/min, at aortic pressure'120mmHg, with heart rate . 120 BPM) for animal implantation of our prototype system.