Abstract
The 8 rare earth elements were determined in the monazite concentrates by X-ray fluorescence spectrometry. The spectral line overlapping interferences from the other elements were investigated and the correction methods of interferences were studied using line overlap coefficients. The coefficients were calculated from the ratios of the intensities measured at the diffracted angle (2${\theta}$) of the analytical lines of other elements to the intensity of pure rare earth oxide. The coefficients were used to correct the line overlaps by a regression analysis. The linearities of calibration curves from the corrected intensities were remarkably improved, and their standard deviations were decreased. The analytical results agreed with that of the inductively coupled plasma(ICP) spectrometry within an allowable error range.
8가지 희토류원소를 주로 포함하고 있는 monazite정광을 X-선 형광분광법으로 분석하는데 있어서 분석선에 대한 다른 희토류원소의 스펙트럼선 겹침 방해를 조사하고 겹침 계수를 구하여 보정하는 방법에 대한 연구를 하였다. 순수한 희토류산화물을 사용하여 원소자신의 분석선 세기와 다른 원소의 분석선에 해당하는 회절각(2${\theta}$)에서의 세기를 측정하고 그들 세기의 비로 겹침계수를 구하였다. 겹침 계수를 이용하여 표준시료와 분석시료의 측정세기를 회귀분석법으로 보정하였다. 보정된 세기를 농도에 대해 도시한 검정곡선은 직선성이 현저히 향상되었고 표준편차도 많이 감소하였다. 시료의 분석 결과도 유도결합플라스마 방출분광법의 결과와 허용오차범위내에서 잘 일치하였다.