The Influence by Stylus Tip Radius and Measuring force on the Stylus Type Surface Roughness Tester

촉침식표면거칠기 측정기에 있어서 촉침의 선단곡률반경과 측정압이 측정에 미치는 영향

  • Published : 1986.03.01

Abstract

The measuring Method by stylus, which measures surface roughness, has been widely used since G. Schmaltz developed the first equipment of that type. Withe the resent development of the measuring method by stylus, surface foughness testing instruments of the very high magnification ratio, hundreds of thousands, are manu- factured. However, as the techniques of the high precision roughness measurement are being required, the response problems due to the tip shapes and the plastic deformations of measured surface of mild material are to be serious factors. In this study, diamond stylus of tip radius $0.5\mu\textrm{m}$, $2\mu\textrm{m}$, $5\mu\textrm{m}$ and $10\mu\textrm{m}$ were used under the measuring force of 0.01gf, 0.02gf, 0.07gf, 0.4gf and 1.6gf, and from the experimental data, maximum measured value devrements between $2\mu\textrm{mR}$ stylus and $5\mu\textrm{mR}$ stylus, $2\mu\textrm{mR}$ stylus and $10\mu\textrm{mR}$ stylus are 22% and 31%, respectively when the measured value of $2\mu\textrm{mR}$ stylus goes to $0.01\mu\textrm{m Ra}$, $\lambda$ c2.5mm. And it is shown that plastic deformations on the plastic deformations on the measured surface are proportional to W/R(W;measured force, R;stylus tip radius).

Keywords