A Study on the Effect of Field Shaping on Dose Distribution of Electron Beams

전자선의 선량분포에 있어서 Field Shaping의 효과에 관한 연구

  • Kang, Wee-Saing (Department of Therapeutic Radiology, College of Medicine, Seoul National University) ;
  • Cho, Moon-June (Department of Therapeutic Radiology, College of Medicine, Seoul National University)
  • 강위생 (서울대학교 의과대학 치료방사선과학교실) ;
  • 조문준 (서울대학교 의과대학 치료방사선과학교실)
  • Published : 1986.12.01

Abstract

In electron therapy, lead cutout or low-melting alloy block is used for shaping the field. Material for shaping electron field affects the output factor as wet 1 as the collimation system. The authors measured the output factors of electron beams for shaped fields from Clinac-18 using ionization chamber of Farmer type in polystyrene phantom. They analyzed the parameters that affect the output factors. The output factors of electron beams depend on the incident energy, collimation system and size of shaped field. For shaped field the variation of output factor for the field size (A/P) has appearence of a smooth curve for all energy and all applicator collimator combination. The output factors for open field deviate from the curves for shaped fields. An output factor for a given field can be calculated by equivalent field method such as A/P method, if a combination of applicator and collimator is fixed.

전자선 치료시에 납판이나 저융점 합금이 조사면의 형태의 변형을 위하여 사용되고 있다. 콜리메이팅 장치와 마찬가지로 조사면 변형을 위한 물질도 전자선의 출력에 영향을 미친다. 저자들은 폴리스티렌 팬톰에 삽입된 Farmer형 전리함을 이용하여 Clinac-18의 전자선의 변형된 조사면에 대한 출력율을 측정하고 출력율에 영향을 미치는 요인들에 관해 분석했다. 전자선의 출력율은 전자선의 초기 에너지, 콜리메이팅 장치뿐만 아니라 조사면 크기에도 좌우되었다. 변형된 조사면에 대하여 잔자선의 에너지에 관계없이 X-선 콜리메이터와 전자선 어플리케이터의 조합이 고정되면 조사면의 크기 대 출력율은 A/P로 표시된 등가조사면의 크기에 따라 변하지만 조사면의 형태에 대해서는 무시할 수 있었다. 그러나 개조사면에 대한 출력율은 변형조사면의 출력율로부터 예상될 수 없고, 그것들만의 독립적인 관계를 가지고 있었다. 어플리케이터와 콜리메이터의 조합이 고정된 경우에 한해서 판자선의 변형조사면에 대한 출력율은 A/P로서 표시된 등가 조사면 방법에 의하여 구할 수 있다.

Keywords